Temperature Advanced Calibrators TA-350P, TA-650P

TA-350P

TA-650P

Key Features :  .

  • – The TA-350P and TA-650P models generate from ambient temperature to 350 ºC and 650 ºC, respectively, with a resolution of 0.01 ºC.
  • – Optional temperature measurement and control by external probe with Callendar-Van Dusen coefficients.
  • – Using the internal reference, accuracy of up to ±(0.1 ºC + 0.1 % of reading).
  • – Using the external probe, accuracy of up to ±0.07 °C for TA-350P and ±0.10 °C for TA-650P.
  • – Using an external thermometer such as Presys Super Thermometer ST-501, accuracy of up to ±0.05 ºC.
  • – Stability of ±0.05 ºC.
  • – Uniformity of ±0.05 ºC.
  • – Input for RTD, thermocouples and thermoswitches.
  • – Internal regulated 24 Vdc power supply and mA input for 2-wire transmitters.
  • – External reference thermometer not necessary.
  • – Completely automatic calibration with or without the use of a computer.
  • – Documenting capabilities: connection with computer and ISOPLAN calibration software.
  • – Portable, compact, with carrying case and interchangeable inserts.

Specification